2 Hz to 85 GHz up to 90 GHz with R&S®FSW-B90G option, YIG preselector = off
Aging of frequency reference
1 × 10–7/year
with R&S®FSW-B4 option
3 × 10–8/year
Bandwidths
Resolution bandwidths
standard filter
1 Hz to 10 MHz, 80 MHz (with R&S®FSW-B8 option)
RRC filter
18 kHz (NADC), 24.3 kHz (TETRA), 3.84 MHz (3GPP)
channel filter
100 Hz to 5 MHz
video filter
1 Hz to 10 MHz
I/Q demodulation bandwidth
10 MHz
with R&S®FSW-B28 option
28 MHz
with R&S®FSW-B40 option
40 MHz
with R&S®FSW-B80 option
80 MHz
with R&S®FSW-B160 option
160 MHz
with R&S®FSW-B320 option
320 MHz
with R&S®FSW-B512 option
512 MHz
with R&S®FSW-B1200 option
1.2 GHz1)
with R&S®FSW-B2001 option
2 GHz1)
with R&S®FSW-B4001 option
4.4 GHz2)
with R&S®FSW-B6001 option
6.4 GHz3)
with R&S®FSW-B8001 option
8.3 GHz3)
with R&S®FSW-B2000 option
2 GHz4)
with R&S®FSW-B5000 option
5 GHz5)
Phase noise
10 kHz offset from carrier
500 MHz carrier
–141 dBc (1 Hz) (typ.)
1 GHz carrier
–140 dBc (1 Hz) (typ.)
10 GHz carrier
–133 dBc (1 Hz) (typ.)
Displayed average noise level (DANL)
2 GHz
–156 dBm (1 Hz) (typ.)
with R&S®FSW-B13 option
–159 dBm (1 Hz) (typ.)
DANL with preamplifier (R&S®FSW-B24 option)
2 GHz
–169 dBm (1 Hz) (typ.)
Intermodulation
Third-order intercept (TOI)
f < 1 GHz
+30 dBm (typ.)
f < 3 GHz
+25 dBm (typ.)
19 GHz to 26.5 GHz
+23 dBm (typ.)
Total measurement uncertainty
8 GHz
< 0.37 dB
1) Not available for the R&S®FSW8 and R&S®FSW13. 2)Available for the R&S®FSW43, R&S®FSW50, R&S®FSW67, R&S®FSW85. 3)Available for the R&S®FSW43, R&S®FSW50, R&S®FSW67, R&S®FSW85 for f ≥ 18 GHz. 4) 2 GHz demodulation bandwidth for frequencies > 5.5 GHz. R&S®RTO2044 digital oscilloscope required. Not available for the R&S®FSW8 and R&S®FSW13. 5) Available for the R&S®FSW85 and R&S®FSW43 (R&S®RTO2064 oscilloscope required).